The INFICON Inverted Magnetron Pirani Gauges, MPG400 and MPG401, measure from 5 x 10-9 mbar to atmosphere (3.8 x 10-9 Torr to atmosphere).
Combining technologies into one single compact unit withone logarithmic analog out put signal significantly reduces the complexity of installation, setup andintegration.
 
| Measurement range (air, N2) | mbar Torr | 5 x 10-9 … 1000 3.8 x 10-9 … 760 | 
| Accuracy | ||
| 1 x 10-8 … 100 mbar | % of reading | ≈±30% | 
| Repeatability | ||
| 1 x 10-8 … 100 mbar | % of reading | ≈±5% | 
| Temperature | ||
| Operation (ambient) | °C | +5 … +55 | 
| Storage | °C | –40 … +65 | 
| Bake-out (1) | °C | 150 | 
| Supply voltage | ||
| At gauge | V (dc) | +15 … +30 | 
| At supply unit with max. cable length (2) | V (dc) | +16 … +30 | 
| Ripple | V (p-p) | ≤1 | 
| Power consumption | W | ≤2 | 
| Materials exposed to vacuum | stainless steel, Al2O3, FPM75, Mo, Ni, Au, W | |
| (1) Without electronics and magnetic shielding. (2) The minimum voltage of the supply unit must be increased proportionally to the length of the sensor cable. | ||