The Rocklabs Standalone RSD (Rotating Sample Divider) is designed to split out a representative portion of the sample with a proportion of 2-50%.
Big inputs and accurate outputs for seriously reliable results.
Benefits:
| Specifications |  | 
| Motor power | 0.18kW, single phase | 
| Maximum input particle size | 10mm | 
| Width x Height x Depth | 1000 x 960 x 800mm | 
| Weight | 135kg | 
| Hopper capacity | 25kg | 
| Dust extraction | 1 extraction port available | 
| Options | Options Sample bins: 
 | 
| Warranty | ROCKLABS Bench Top RSD is sold with a 2000-operating-hours or one year warranty (whichever comes first), covering all parts (excluding wear parts) and labour if used under laboratory conditions. | 
| Comments on taking a representative split | Taking a representative split from a large sample is not as simple as many people believe. Rolling on a mat, multiple grabs or riffle splitting do not produce accurate results. A classic study* conducted at Bradford University demonstrated that a rotating sample divider provided the best results. * Critical evaluation of powder sample procedures. Allen T and Khan A. A, The chemical Engineer, May 1979, pp108 - 112 | 
| Dividing | Standard deviation of samples % | 
| Cone and quartering | 6.810 | 
| Scoop sampling | 5.140 | 
| Riffle splitting | 1.010 | 
| Table sampling | 2.090 | 
| Rotating splitting | 0.125 | 
| Random Variation | 0.076 | 
 Malaysia