The Rocklabs Benchtop RSD (Rotating Sample Divider) is designed to be placed on a work bench and can be used in conjunction with Benchtop Ring Mill.
Big inputs and accurate outputs for seriously reliable results.
Benefits:
| Specifications |  | 
| Motor power | 0.1kW, single phase | 
| Maximum input particle size | 10mm | 
| Width x Height x Depth | 600 x 450.2 x 420mm | 
| Weight | 58.82kg | 
| Hopper capacity | 5kg | 
| Dust extraction | n/a | 
| Options | n/a | 
| Warranty | ROCKLABS Bench Top RSD is sold with a 2000-operating-hours or one year warranty (whichever comes first), covering all parts (excluding wear parts) and labour if used under laboratory conditions. | 
| Comments on taking a representative split | Taking a representative split from a large sample is not as simple as many people believe. Rolling on a mat, multiple grabs or riffle splitting do not produce accurate results. A classic study* conducted at Bradford University demonstrated that a rotating sample divider provided the best results. * Critical evaluation of powder sample procedures. Allen T and Khan A. A, The chemical Engineer, May 1979, pp108 - 112 | 
| Dividing | Standard deviation of samples % | 
| Cone and quartering | 6.810 | 
| Scoop sampling | 5.140 | 
| Riffle splitting | 1.010 | 
| Table sampling | 2.090 | 
| Rotating splitting | 0.125 | 
| Random Variation | 0.076 | 