Force in 3 dimensions(X, Y, Z) can be measured at the same time
Safe overload rating of 500%
±5 V output at the rated capacity for each direction: Fx, Fy, Fz (three axis).
As output signals for each direction is adjusted before shipping out, you can get calibrated(scaled) output by just connecting sensor and DC 24 V power supply to FS300.

 
| Model | Sensor:UPP-100NX3, Amplifier:FS300 | 
|---|---|
| Rated capacity | Fx, Fy : ±100 N, Fz : 100 N | 
| Power supply voltage | DC 24 V±15% | 
| Consumption current | 150 mA or less | 
| Load signal output | Fx, Fy, Fz:±5 V | 
| Load resistance | 2 k!& or more | 
| Response | 2 kHz | 
| I/O signal | Input: Auto zero command Output: Auto zero response  | 
		
| Safe overload | 500% FS | 
| Non-linearity | 0.5% FS or less | 
| Hysteresis | 0.5% FS or less | 
| Repeatability | 0.5% FS or less | 
| Compensated temperature range  | 
			0 to +40℃ | 
| Safe temperature range | -10 to +50℃ | 
| Temperature effect on zero | 0.5% FS/℃ or below | 
| Temperature effect on span | 0.05% FS/℃ or below | 
| Sensor material | Aluminum alloy | 
| Weight | Sensor:Approx. 70 g, Amplifier:Approx. 180 g  | 
		
| Cable | 8 conductor flexible cable 1 m (Connector is included for FS300 connection)  | 
		
* Above specifications are achieved when the sensor(UPP-100NX3) is connected with the dedicated amplifier(FS300).
* Accuracy is not guaranteed in the range of Fz 0 to -5 V.
 

Please note that there are possibilities of individual differences in a color tone on display devices such as LEDs, fluorescent display tubes and LCDs due to manufacturing process or production lots.
 