Comet T3118 Specifications |
|
Technical parameters |
Value |
|
Measured Value |
Voltage |
|
Operating temperature range |
-30 to +70°C |
|
Measurement accuracy |
±0.2% FS |
|
Resolution of voltage input |
13 bits (8192 levels) |
|
Signal character at binary input |
from potential-less contact or two-state voltage signal |
|
Minimum pulse duration at binary input |
500 ms (shorter pulses may not be recorded) |
|
Maximum frequency at binary input |
0.5 Hz (i.e. maximum 5 pulses for 10 s) |
|
Power current through contact at binary input |
3 uA (contact closed) |
|
Voltage accross open contact at binary input |
maximum 3.6 V |
|
Low voltage level at binary input |
0 to +0.2 V (maximum current from the input 3 uA |
|
High voltage level at binary input |
+3.0 to +30 V (maximum current to the input 100 nA |
|
Built-in connector for input signals |
D-Sub 9 pins, male |
|
Real time clock |
year, leap year, month, day, hour, minute, second |
|
Data logging interval |
adjustable from 10s to 24h |
|
Display and alarm refresh |
each 10s |
|
Total memory capacity |
32 504 records in noncyclic logging mode, 28 896 records in cyclic logging mode |
|
Data logging modes |
noncyclic - data logging stops after filling the memory cyclic - after filling memory oldest data is overwritten by new |
|
Power |
Lithium battery 3,6V, size AA |
|
Typical battery life in low-power mode (1 minute sampling) |
6 years |
|
Battery life in continuous on-line mode with interval 1 minute |
4 years |
|
Battery life in continuous on-line mode with interval 10 second |
1 year |
|
IP protection |
IP67 - protected against influence of temporary immersion into water |
|
Dimensions without connectors |
93x64x29mm |
|
Weight including battery |
approximately 130g |