Salt Spray Corrosion Test Chamber
A professional-grade environmental simulation system designed to evaluate the anti-erosion resistance of treated material surfaces. It provides critical data for products treated with electric plating, anodizing, and organic coatings through standardized saline and corrosive mist exposure.
Product Overview
Technical Highlights
- Internal Volume110L to 1500L Capacity
- Lab Temp (NSS/ACSS)35±1°C
- Lab Temp (CASS)50±1°C
- Spray Volume1.0-2.0ml / 80cm² / h
- NaCl Concentration5% Standardized Solution
Testing Parameters
- Air Pressure1.0 ~ 6.0 Kgf
- Bucket Temp (NSS)47±1°C
- Bucket Temp (CASS)63±1°C
- Power SupplyAC 220V 15A
- Spray FormProgrammable / Continuous
Performance Advantages
Precision Atomization
Achieves a standardized spray volume of 1.0-2.0ml/80cm²/h, ensuring uniform distribution of corrosive mist across all samples.
Flexible Programming
Supports both continuous and programmable spraying cycles to accommodate specific research requirements and international protocols.
Dual-Mode Thermal Control
Independent temperature stabilization for the laboratory chamber and the pressure bucket to support NSS, ACSS, and CASS testing methods.
Scalable Capacity
Available in a wide range of internal dimensions from 600mm to 2000mm in width to test components of varying sizes and quantities.
Core Systems
Technical Specifications
| Parameter / Model Breakdowns | SH-60 | SH-90 | SH-120 | SH-160 | SH-200 |
|---|---|---|---|---|---|
| Internal Dimension WxHxD (mm) | 600x450x400 | 900x600x500 | 1200x800x500 | 1600x800x500 | 2000x1000x500 |
| External Dimension WxHxD (mm) | 1100x600x1200 | 1400x950x1200 | 1900x1300x1400 | 2300x1300x1400 | 2700x1500x1500 |
| Lab Capacity (L) | 110 | 270 | 500 | 800 | 1500 |
| Saline Capacity (L) | 15 | 25 | 40 | 40 | 40 |
| Lab Temperature (NSS ACSS) | 35±1 °C | ||||
| Lab Temperature (CASS) | 50±1 °C | ||||
| Pressure Bucket Temperature (NSS ACSS) | 47±1 °C | ||||
| Pressure Bucket Temperature (CASS) | 63±1 °C | ||||
| Saline Concentration | 5% concentration of NaCl, or every Liter 5% NaCl add 0.26g CuCl 侶 侽 | ||||
| Compressed Air Pressure (Kgf) | 1.0~6.0 | ||||
| Spray Volume Rate | 1.0~2.0ml /80cm² /h (working at least 16 hours, and take the average value) | ||||
| Spray Form | Continuous Spraying or Programmable Spraying | ||||
| Power Requirement | AC 220V, 1φ 3 lines 15A | ||||
| Weight / Mass Range (Kg) | 65 | 80 | 120 | 160 | 200 |
Primary Use Cases
Surface Treatment
Testing endurance of electric plating, anodizing, and spray coating on metallic components.
Quality Assurance
Verification of anti-rusting oils and chemical anti-erosion treatments for industrial hardware.
Automotive & Aerospace
Accelerated corrosion testing for exterior parts exposed to harsh salt and moisture environments.