The Elcometer 6910 Setaflash ‘Series 8’ Closed Cup Tester is an automated closed cup tester with a temperature range of ambient to 300°C (ambient to 572°F).
Test parameters, instrument status and flash point results are shown on the bright colour LCD display – all easily accessed via menus, navigated by the keypad, which guides the operator through a straightforward testing process.
The instrument uses an electric hot wire ignitor and is designed with two modes of flash point operation; Flash/No Flash and Ramp. Ramp Mode increases the sample temperature at the rate of 2°C/min and automatically dips the ignitor every 1°C up to 100°C (every 2°C above 100°C), until either a flash is detected or the end of test temperature is reached.
During the test the automatic flash point is set to ignore halo and other false flash effects reducing the risk of inaccurate test results and enhancing repeatability. At the end of the flash point test, the sample cup is cooled by forced air to reduce the cycle time between tests.
Up to 64 results can be stored in the memory which can be downloaded to a PC or printer via the USB port for further reporting.
| Part Number | Description | Certificate |
| K6910M020 | Elcometer 6910 Setaflash ‘Series 8’ Closed Cup Tester | ![]() |
| Temperature Range | Ambient to 300°C (Ambient to 572°C) | |
| Sample Size | 2ml (0.067fl oz) or 4ml (0.135fl oz) according to method | |
| Test Modes | Rapid Equilibrium and Ramp | |
| Test Duration Rapid Equilibrium Mode | 1 minute below 100°C (212°F) 2 minutes above 100°C (212°F) |
|
| Test Duration Ramp Mode | Typically 7 minutes | |
| Cool Down Time | 80°C to 20°C (176°F to 68°F) in 4 minutes | |
| Cup Material | Aluminium | |
| Dimensions | 300 x 340 x 380mm (11.8 x 13.4 x 15.0”) | |
| Weight | 8kg (17.6lb) | |
| Packing List | Elcometer 6910 Setaflash ‘Series 8’ Closed Cup Tester, mains power cable, 2ml (0.067fl oz) syringe, programming plug, 2mm Allen key, 2.5mm Allen key, ‘O’ ring fitted, certified flash point material (50ml/1.7fl oz), quality assurance certificate and operating instructions | |