Compatible to the wide range of INFICON active gauges, the new VGC50x series of active gauge controllers are able to monitor and data log the entire pressure range from 10 {¹ p to 1500 mbar (10 {¹ p to 1125 Torr) and the set point status.
| Type | VGC501 | |
| Measurement channels | 1 | |
| Display | LCD, LED backlight | |
| Rate | 1/s | 10 |
| Connectable gauges with display range | ||
| CDG (A/D) | Torr | 1×10-3×F.S. …1×F.S. |
| PCG | mbarTorr | 5×10-4…1500 |
| PSG | mbarTorr | 5×10-4…1000 |
| MPG | mbarTorr | 1×10-9…1000 |
| PEG | mbarTorr | 1×10-9…1×10-2 |
| MAG | mbarTorr | 1×10-9…1×10-2 |
| BCG | mbarTorr | 5×10-10…1500 |
| BPG | mbarTorr | 5×10-10…1000 |
| HPG | mbarTorr | 2×10-6…1000 |
| Measurement unit (selectable) | mbar, Torr, Pa, hPa, Micron, V | |
| Setpoints | ||
| Setpoint relays | Setpoint relays | 2 |
| Channel assignment | Channel assignment | 1 |
| Adjustment range | Adjustment range | sensor dependent |
| Hysteresis | Hysteresis | adjustable |
| Relay contact | Relay contact | floating changeover contact |
| Contact rating | V (ac) • AV (dc) • A | 30 / 1 |
| Connector | Connector | D-sub, 15-pin, male |
| Analog output | ||
| Range | V | 0 ... 10.3, sensor analog output signal |
| Analog output | Analog output | 1 |
| Programmable analog output | Programmable analog output | - |
| Connector | Connector | D-Sub, 15-pin, male |
| Interface (digital) | ||
| Connector | Connector | USB slave, master and Ethernet, USB Typ A (stick), USB Type B, FCC68/RJ45 |
| Power | ||
| Supply | V (ac) | 100 ... 240 |
| Frequency | Hz | 50 ... 60 |
| Consumption | W | ≤45 |
| Operation temperature (ambiance) | °C | +5 ... +50 |