※imageWhen the detector on the top of the unit is brought close and AC voltage is detected, a buzzer and LED will sound to notify you.
| range | resolution | Accuracy | remarks | |
|---|---|---|---|---|
| DCV | 600 mV / 6 V / 60 V / 600 V | 0.1 mV /0.001 V / 0.01 V / 0.1 V | ±(0.5% + 2 dgt) |
|
| 1000 V | 1 V | ±(0.8% + 3 dgt) | ||
| ACV | 600 mV / 6 V / 60 V | 0.1 mV /0.001 V / 0.01 V | ±(0.8% + 2 dgt) |
|
| 600V / 750V | 0.1V / 1V | ±(1.0% + 3 dgt) | ||
| DCA | 600 μA / 6000 μA / 60 mA / 600 mA | 0.1 μA / 1 μA / 0.01 mA / 0.1 mA / | ±(1.0%rdg + 2dgt) |
|
| 10A | 0.01 A | ±(1.5%rdg + 3dgt) | ||
| ACA | 600 μA / 6000 μA / 60 mA / 600 mA | 0.1 μA / 1 μA / 0.01 mA / 0.1 mA | ±(1.0% + 3 dgt) |
|
| 10A | 0.01 A | ±(1.5% + 3 dgt) | ||
| Ω | 600Ω / 6kΩ / 60kΩ / 600kΩ | 0.1 Ω / 0.001 kΩ / 0.01 kΩ / 0.1 kΩ | ±(0.8%rdg + 2dgt) |
|
| 6 MΩ / 60 MΩ | 0.001 MΩ / 0.01 MΩ | ±(2.0%rdg + 3dgt) | ||
| capacitance | 60nF / 600nF / 6μF / 60μF | 0.01 nF / 0.1 nF / 0.001 μF / 0.01 μF | ±(3.0%rdg + 3dgt) |
|
| 600 μF / 6 mF / 60 mF | 0.1 μF / 0.001 mF / 0.01 mF | ±(5.0%rdg + 5dgt) | ||
| temperature | 1000℃ | 1°C | ±(2.5%rdg + 3dgt) |
|
| Buzzer Threshold | remarks | |
|---|---|---|
| Continuity check | Approximately 50 MΩ |
|
| range | resolution | remarks | |
|---|---|---|---|
| Diode Check | 0 V to 3 V | 1 mV |
|
| Measurement range | Accuracy | frequency | |
|---|---|---|---|
| frequency | 1 Hz to 5 MHz | ±(0.8%rdg + 2dgt) | 1.0 Vrms or more, sine wave, 50% square wave |
| DUTY | 1.0 to 99.0% | ±(2.5% + 3 dgt) | 10 Hz to 10 kHz, 1.5Vrms or more, square wave |
| Voltage tester | AC 90 V to AC 1000 V | - | 50 Hz or 60 Hz |
| Detection method | True RMS |
|---|---|
| display | LCD, maximum display “6000” |
| Range switching | Auto/Manual |
| polarity display | Automatic switching, only "-" displayed |
| sampling | Approximately 3 times/second |
| Auto Power Off | Approximately 30 minutes (can be canceled) |
| Operating temperature and humidity | 0 to +40°C, 80% RH or less (no condensation) |
| Storage temperature and humidity | -10 to +60°C, 80% RH or less (no condensation) |
| size | Approx. W 74 × H 148 × D 38 mm |
| weight | Approximately 210 g (including batteries) |
| Test lead length | Approximately 75cm |
| power supply | 2 x AA batteries |
| battery life | Approximately 900 hours (when using alkaline batteries) |
| accessories | Instruction manual, test leads, K-type thermocouple, battery cover screwdriver |
| option | |
| proofreading | Possible |
| Compliance standards | IEC61010-1 CAT III 600V |
| JAN | 4983621100433 |