Detection method |
Absorbance |
|
Detection mode |
Endpoint, kinetic |
|
Microplates types |
96-well microplates |
|
Light source |
Up to 4 LEDs |
|
Detector |
96 Photodiodes |
|
Wavelength selection |
Example combinations (FWHM 10 nm): 450, 492, 562, 620 nm / 405, 450, 540, 630 nm / 492, 562, 605, 650 nm. Other filter combinations in the range of 400–1000 nm possible. |
|
Photometric range |
0.0–4.0 OD |
|
Linearity |
405 nm: ≤ 1.5 % from 0.0–2.0 OD; ≤ 3.0 % from 2.0–3.0 OD≥ 450 nm: ≤ 1.0 % from 0.0–2.0 OD; ≤ 1.5 % from 2.0–3.0 OD |
|
Accuracy |
405 nm: ≤ 1.5 % + 0.010 OD from 0.0–2.0 OD; ≤ 3.0 % + 0.010 OD from 2.0–3.0 OD≥ 450 nm: ≤ 1.0 % + 0.010 OD from 0.0–2.0 OD; ≤ 1.5 % + 0.010 OD from 2.0–3.0 OD |
|
Reproducibility |
≤ 0.5 % + 0.005 OD from 0.0–2.0 OD; ≤ 1 % + 0.010 OD from 2.0–3.0 OD |
|
Resolution |
0.001 OD |
|
Read time |
Dow to 3 sec. at single wavelength |
|
Data output |
USB 2.0 interface with PC |
|
Dimensions |
9.6 x 15.4 x 5.5 cm (W x L x H) |
|
Weight |
900 g |
|
Power supply |
USB connection 5 V |
|
Power consumption |
2.5 Watts |
|
Operating systems |
Microsoft Windows, macOS |
|
Measurement modes |
Endpoint, Kinetic |
|
Evaluation methods |
Quantitative, Qualitative |
|
Regression methods |
Point-to-Point, Linear, Four Parameter Logistic (4PL), Five Parameter Logistic (5PL) |
|
Export formats |
PDF, CSV |