Model No. (Order Code)
Note: MR8848 cannot operate alone. You need to purchase and install one or more optional input modules, etc.
FEATURES
For Evaluation Testing and Troubleshooting
Robust Housing
To minimize the risk of failure due to drops and shocks, the Memory HiCorder MR8848 employs a rugged housing with protectors for enhanced durability.
Safe and Reliable Measurements
When inspecting lines of differing electrical systems simultaneously (e.g., power lines and signals), short-circuit are a serious risk. Unlike oscilloscopes, memory HiCorders have isolated channels, so short-circuit accidents will not occur between differing circuits.
Correlation Analysis of Various Physical Phenomena
Troubleshooting and analysis can be more in depth than ever before through simultaneous recording of various physical phenomena.
The Memory HiCorder MR8848 records and analyzes multiple physical parameters—including voltage, current, temperature, vibration, pressure, and strain—for identifying correlations and root causes.
Onsite, Tamper-Proof Printing (Factory-Installed Option)
The factory-option printer lets you immediately print and share tamper-proof hard copies of your measurements right at the site, so that data cannot be tempered with.
Record Long-Duration Waveforms with High-Capacity Storage
Long-term waveform recording becomes possible by storing data on the large-capacity 1 TB SSD (U8334) with the Direct Write to Storage option (MR9001-01).
For example, in railway vehicle measurement, the Memory HiCorder will record over 10 parameters at a sampling rate of 10 µs/S for more than an hour.
SPECIFICATIONS
Basic specifications
Accuracy guaranteed: 1 year
| Measurement functions |
Memory function (waveform recording), Recorder function (peak/trough recording), X-Y Recorder function, FFT function |
| Max. number of channels |
16 ch analog + 16 ch logic
32 ch analog + 16 ch logic (when U8975 or U8978 × 8 is attached)
10 ch analog + 64 ch logic (when main unit logic + Logic Unit 8973 × 3 is attached)
20 ch analog + 64 ch logic (when U8975 or U8978 × 5 + main unit logic + Logic Unit 8973 × 3 is attached) |
| Module slots |
Up to 8 modules
Restrictions:
Current Unit 8971: up to 4 modules can be installed.
Logic Unit 8973: up to 3 modules can be installed.
3CH Current Unit U8977: up to 3 modules can be installed. |
| Number of main unit logic channels |
16 (The GND of the logic probe input connector is common with the main unit GND)
When the Digital Voltmeter Unit MR8990 is installed in Unit 1 and 2, the main unit logic cannot be used.
Following restrictions when using the main unit logic (when logic measurement is turned ON):
- The measurement resolution of the modules installed in Unit 1 and 2 will be limited to 12 bits.
- Frequency modules installed in Unit 1 and 2 cannot be used. |
| Max. sampling speed |
20 MS/s for simultaneous use of all channels (when using the Analog Unit 8966)
10 MS/s for external sampling |
| Memory capacity |
Total 512 megawords (no memory expansion)
(16 megawords/ch using 32 analog channels, 256 megawords/ch using 2 analog channels) |
| Power supply |
100 V to 240 V AC, 50/60 Hz, 10 V to 28 V DC (when using the DC Power Unit 9784) |
| Max. rated power |
130 VA (220 VA when using the Printer Unit U8351) |
| External interfaces |
LAN, USB, external control terminals |
| External storage |
SD card slot, built-in drive (1 TB, factory-installed option), USB flash drive |
| Printer |
When the factory-installed option Printer Unit U8351 is installed.
Thermal recording method, use the Recording Paper 9231, recording speed 50 mm/s |
| Other functions |
Trigger, numerical calculation, waveform calculation, storage partitioning, cursor measurement, monitor, position display (VIEW), waveform evaluation, waveform generation |
| Dimensions and weight |
Approx. 351W × 261H × 140D mm (13.8W × 10.3H × 5.5D in.) (excluding protrusions)
Approx. 365W × 307H × 160D mm (14.4W × 12.1H × 6.3D in.) (including protrusions)
Approx. 6.9 kg (15.2 lbs.) (main unit only)
Approx. 7.4 kg (16.3 lbs.) (when the Printer Unit U8351 is installed) |
| Included accessories |
Startup Guide × 1, Operating Precautions × 1, input cord label × 1, power cord × 1, Recording Paper 9231 (when Printer Unit U8351 is installed) × 1, roll paper attachment (when Printer Unit U8351 is installed) × 1 |
OPTIONS
Factory-installed option (3)
Please specify these requirements when placing your order, as it will be built in during production.
INTERNAL STORAGE U8334
1 TB, FAT32, 1 year product warranty
PRINTER UNIT U8351
Thermal recording method, 50 mm/s max. recording speed
DC POWER UNIT 9784
Factory-installed option (not user-installable, built in on the bottom casing, 10 to 28 V DC)
Additional function (1)
DIRECT WRITE TO STORAGE MR9001-01
Function to record measurement waveforms directly to storage devices (one-time purchase of license key)
Input modules (14)
Input cables are not included. Please purchase them separately.
DIGITAL VOLTMETER UNIT MR8990
0 For minute voltage measurement
0 Max. 0.1 μV resolution
0 500 times/s sampling
STRAIN UNIT U8969
• 2 ch, acceleration, load, pressure or strain
• Strain gauge converter amp
HIGH VOLTAGE UNIT U8974
0 2 ch, high-voltage input
0 Max. 1000 V DC, 700 V AC rms
0 DC to 100 kHz bandwidth
4CH ANALOG UNIT U8975
0 4 ch
0 DC to 2 MHz bandwidth
3CH CURRENT UNIT U8977
0 3 ch, current measurement
0 DC to 2 MHz bandwidth
4CH ANALOG UNIT U8978
0 4 ch instantaneous waveforms
0 DC to 2 MHz bandwidth
CHARGE UNIT U8979
0 Measure acceleration waveforms across 2 channels
0 Charge output type acceleration pickup
0 Preamplifier built-in acceleration pickup
ANALOG UNIT 8966
• 2 ch, instantaneous waveforms
• DC to 5 MHz bandwidth
TEMP UNIT 8967
• 2 ch, temperature
• Thermocouple input
HIGH RESOLUTION UNIT 8968
• 2 ch, high resolution measurement
• DC to 100 kHz bandwidth
FREQ UNIT 8970
• 2 ch, rotation, frequency
• DC to 100 kHz
CURRENT UNIT 8971
• 2 ch, measuring current
• DC to 100 kHz bandwidth
• Directly connectable with current sensors
DC/RMS UNIT 8972
• 2 ch, instantaneous waveform/RMS
• DC to 400 kHz bandwidth
• RMS: DC and 30 to 100 kHz
LOGIC UNIT 8973
• 16 ch, logic signal
• Up to four logic probes
Output modules (3)
Output cables not included. Please purchase them separately.
WAVEFORM GENERATOR UNIT MR8790
01 Hz to 20 kHz sine waveform output
0±10 V DC output
04 channels
PULSE GENERATOR UNIT MR8791
00.1 Hz to 20 kHz pulse
0Pattern output
08 channels
ARBITRARY WAVEFORM GENERATOR UNIT U8793
010 mHz - 100 kHz FG
0Max. 15 V output
02 channels
Carrying case (1)
CARRYING CASE 9783
Equipped with an options compartment, durable hard-type trunk
PC Software (4)
FlexPro
• Third-party software
• For detailed analysis of Memory HiCorder data
LAN COMMUNICATOR 9333
• For LAN communication with the Memory HiCorder
• Data collection and remote control
WAVE PROCESSOR 9335
• For the Memory HiCorder
• Convert data, print and display waveforms
LAN CABLE 9642
Straight Ethernet cable, supplied with straight-to-cross conversion adapter, 5 m (16.4 ft.)
Storage media (3)
Please be sure to use Hioki's optional storage media. If you use storage media made by other manufacturers, saving or reading data may not function properly, and we cannot guarantee the compatibility and performance.
SD MEMORY CARD 2GB Z4001
2 GB capacity
SD MEMORY CARD Z4003
8 GB capacity
USB DRIVE Z4006
16 GB, long-life, high-reliability SLC flash memory
Logic probe (3)
LOGIC PROBE MR9321-01
• 4 isolated channels
• ON/OFF detection of AC/DC voltage
• Large terminal type
LOGIC PROBE 9320-01
• 4 channels for contact voltage signals
• 500 ns response
• Miniature terminal type
LOGIC PROBE 9327
• 4 channels for contact voltage signals
• High-speed 100 ns response
Recording paper (1)
For Printer Unit U8351
RECORDING PAPER 9231
Roll type A4 width 216 mm (8.50 in) × 30 m (98.43 ft), 6 rolls/set
Output cable (2)
CONNECTION CABLE L9795-01
Max. rated voltage to earth: 33 Vrms AC or 70 V DC, SMB to alligator clip, 1.5 m (4.92 ft) length
CONNECTION CABLE L9795-02
Max. rated voltage to earth: 33 Vrms AC or 70 V DC, SMB to BNC terminal, 1.5 m (4.92 ft) length
Input cable A (4)
Input voltage is limited to the specifications of the input modules in use.
CONNECTION CORD L9790
• Flexible φ 4.1 mm (0.16 in) thin dia.
• Cable allowing for up to 600 V input
• 1.8 m (5.91 ft) length
ALLIGATOR CLIP L9790-01
Red/black set attaches to the ends of the Connection Cord L9790
GRABBER CLIP 9790-02
Red/black set attaches to the ends of the Connection Cord L9790
CONTACT PIN 9790-03
Red/black set attaches to the ends of the Connection Cord L9790
Input cable B (3)
Input voltage is limited to the specifications of the input modules in use.
CONNECTION CORD L9197
•φ 5.0 mm (0.20 in) dia
•Cable allows for up to 600 V input
•1.8 m (5.91 ft) length
•Includes detachable large alligator clips
CONNECTION CORD L9198
•φ 5.0 mm (0.20 in) dia.
•Cable allows for up to 300 V input
•1.7 m (5.58 ft) length
•Small alligator clip
GRABBER CLIP L9243
Attaches to the tip of the banana plug cable, Red/Black: 1 each, 185 mm (7.28 in.) length, CAT II 1000 V
Input cable C (2)
The maximum input voltage is derated depending on the input frequency. For details, please refer to the instruction manual of each probe.
10:1PROBE 9665
01 kV rms Max. (500 kHz or less)
0Max. rated voltage to earth depends on input unit.
100:1PROBE 9666
05 kV peak Max. (1 MHz or less)
0Max. rated voltage to earth depends on input unit.
Input cable D (2)
The voltage to ground must be within the specified specification of the product. Power supply is required separately.
DIFFERENTIAL PROBE P9000
• Waveform only for exclusive use with Memory HiCorders (P9000-01) or waveform/RMS value switchable (P9000-02)
• For use with input of up to 1 kV AC/DC
• Requires separate power supply
AC ADAPTER Z1008
100 V to 240 V AC
Input cable E (2)
The voltage to ground must be within the specified specification of the product. Power supply is required separately.
DIFFERENTIAL PROBE 9322
• For up to 2 kV DC or 1 kV AC
• Frequency bandwidth up to 10MHz
AC ADAPTER 9418-15
100 to 240V AC
Input cable F (6)
For banana terminals. Input voltage is limited to the specifications of the input modules in use.
EXTENSION CABLE SET L4931
Expands the length of L4930/4940, 1.5 m (4.92 ft) length
ALLIGATOR CLIP SET L4935
Attaches to the tip of the L4930/4940, CAT IV 600 V, CAT III 1000 V
BUS BAR CLIP SET L4936
Attaches to the tip of the L4930/4940, CAT III 600 V
MAGNETIC ADAPTER SET L4937
Attaches to the tip of the L4930/4940, CAT III 1000 V
CONNECTION CABLE SET L4940
Banana plug - banana plug, 1.5 m (4.92 ft) length, red/black each 1
GRABBER CLIP L9243
Attaches to the tip of the banana plug cable, Red/Black: 1 each, 185 mm (7.28 in.) length, CAT II 1000 V
Input cable G (1)
For MR8990. Input voltage is limited to the specifications of the input modules in use.
TEST LEAD L2200
70 cm (2.30ft) length, detachable large alligator clips or needle tips are bundled, CAT IV 600V, CAT III 1000V
Input cable H (1)
For U8979.
CONNECTION CORD 9166
Metal BNC to clip, 1.5 m (4.92 ft) length
Other options for input (2)
CONNECTION CORD L9217
Cord has insulated BNC connectors at both ends, 1.6 m (5.25 ft) length
CONVERSION ADAPTER 9199
Receiving side banana (female), output BNC (male)
Current measurement (high-accuracy) (19)
ME15W (12pin) terminal type
Power supply is required separately to use the current sensors.
AC/DC CURRENT PROBE CT6830
• Band: DC to 100 kHz
• Rated measurement current: 2 A AC/DC
• Measurable conductor: # 5 mm
• For power analyzers and 3 ch current unit of Memory HiCORDER
AC/DC CURRENT PROBE CT6831
• Band: DC to 100 kHz
• Rated measurement current: 20 A AC/DC
• Measurable conductor: # 5 mm
• For power analyzers and 3 ch current unit of Memory HiCORDER
AC/DC CURRENT PROBE CT6833, CT6833-01
• DC to 50 kHz
• Rated 200 A AC/DC
• For vehicle power testing, compliant with WLTP & SAE J1634 standards
AC/DC CURRENT PROBE CT6834, CT6834-01
• DC to 50 kHz
• Rated 500 A AC/DC
• For vehicle power testing, compliant with WLTP & SAE J1634 standards
AC/DC CURRENT PROBE CT6841A
• DC to 2 MHz
• Rated 20 A AC/DC
• Automatic phase correction function
AC/DC CURRENT PROBE CT6843A
• DC to 700 kHz
• Rated 200 A AC/DC
• Automatic phase correction function
AC/DC CURRENT PROBE CT6844A
• DC to 500 kHz
• Rated 500 A AC/DC
• Automatic phase correction function
AC/DC CURRENT PROBE CT6845A
• DC to 200 kHz
• Rated 500 A AC/DC
• Automatic Phase Correction function
AC/DC CURRENT PROBE CT6846A
• DC to 100 kHz
• Rated 1000 A AC/DC
• Automatic phase correction function
AC/DC CURRENT SENSOR CT6862-05
• DC to 1MHz, ±0.05% accuracy
• Rated 50A AC/DC
AC/DC CURRENT SENSOR CT6863-05
• DC to 500kHz, ±0.05% accuracy
• Rated 200A AC/DC
AC/DC CURRENT SENSOR CT6872
• DC to 10 MHz
• Rated 50 A AC/DC
AC/DC CURRENT SENSOR CT6873
• DC to 10 MHz
• Rated 200 A AC/DC
AC/DC CURRENT SENSOR CT6875A
• DC to 2 MHz
• Rated 500 A AC/DC
AC/DC CURRENT SENSOR CT6876A
• DC to 1.5 MHz
• Rated 1000 A AC/DC
AC/DC CURRENT SENSOR CT6877A
• DC to 1 MHz
• Rated 2000 A AC/DC
AC/DC CURRENT SENSOR CT6904A, CT6904A-1
• DC to 4 MHz (CT6904A), DC to 2 MHz (CT6904A-1)
• Rated 500 A AC/DC
AC/DC CURRENT SENSOR CT6904A-2, CT6904A-3
• DC to 4 MHz (Model CT6904A-2), DC to 2 MHz (Model CT6904A-3)
• Rated 800 A AC/DC
CLAMP ON SENSOR 9272-05
• 1 Hz to 100 kHz, ±0.3% accuracy
• Rated 20 A/200 A AC
Power supply (for high-accuracy current) (4)
SENSOR UNIT CT9555
• 1 channel
• Waveform output
SENSOR UNIT CT9556
• 1 channel, ±0.2% DC accuracy
• Waveform and RMS output
SENSOR UNIT CT9557
• 4 channels, ±0.06% DC accuracy
• Waveform, total waveform, and total RMS output
CONNECTION CORD L9217
Cord has insulated BNC connectors at both ends, 1.6 m (5.25 ft) length
Conversion cable (1)
CONVERSION CABLE CT9900
Convert PL23 (10-pin) terminal to ME15W (12-pin) terminal
Current measurement (general use) (8)
PL14 terminal type
Conversion Cable CT9920 is required separately to use the current sensors.
AC FLEXIBLE CURRENT SENSOR CT7040 series
• 10 Hz to 50 kHz, ±1.5% accuracy
• 600A or 6000A AC rated
AC/DC CURRENT SENSOR CT7631
• DC to 10 kHz, ±1.0% accuracy
• Rated 100 A AC/DC
AC/DC CURRENT SENSOR CT7636
• DC to 10kHz, ±2.0% accuracy
• Rated 600 A AC/DC
AC/DC CURRENT SENSOR CT7642
• DC to 10kHz, ±1.5% accuracy
• Rated 2000 A AC/DC
AC/DC AUTO-ZERO CURRENT SENSOR CT7731
• DC to 5 kHz, ±1.0% accuracy
• Rated 100A AC/DC
AC/DC AUTO-ZERO CURRENT SENSOR CT7736
• DC to 5 kHz, ±2.0% accuracy
• Rated 600A AC/DC
AC/DC AUTO-ZERO CURRENT SENSOR CT7742
• DC to 5 kHz, ±1.5% accuracy
• Rated 2000A AC/DC
CONVERSION CABLE CT9920
Required to connect the PW3390 or other instrument’s ME15W terminal to a current sensor with a PL14 output connector.
Current measurement (high-sensitivity, wide-bandwidth) (8)
Power supply is required separately to use the current probes.
CURRENT PROBE CT6700
• DC to 50 MHz (−3dB), ±3.0% accuracy
• Maximum 5 A rms
• ±7.5 A peak (non-continuous)
CURRENT PROBE CT6701
• DC to 120 MHz (−3dB), ±3.0% accuracy
• Maximum 5 A rms
• ±7.5 A peak (non-continuous)
CURRENT PROBE CT6710
• DC to 50 MHz (−3dB), ±3.0% accuracy
• Maximum 30 A rms, 3 ranges
• ±50 A peak (non-continuous)
CURRENT PROBE CT6711
• DC to 120 MHz (−3dB), ±3.0% accuracy
• Maximum 30 A rms, 3 ranges
• ±50 A peak (non-continuous)
CLAMP ON PROBE 3273-50
• DC to 50MHz (−3dB), ±1.0% accuracy
• Maximum 30 A rms
• ±50 A peak (non-continuous)
CLAMP ON PROBE 3274
• DC to 10MHz (−3dB), ±1.0% accuracy
• Maximum 150 A rms
• ±300 A peak (non-continuous)
CLAMP ON PROBE 3275
• DC to 2MHz (−3dB), ±1.0% accuracy
• Maximum 500 A rms
• ±700 A peak (non-continuous)
CLAMP ON PROBE 3276
• DC to 100 MHz (−3dB), ±1.0% accuracy
• Maximum 30 A rms
• ±50 A peak (non-continuous)
Power supply (for high-sensitivity, wide-bandwidth current) (2)
POWER SUPPLY 3269
• 2 channels for CT6710 series or 4 channels for CT6700 and 3270 series
• ±12 V output
POWER SUPPLY 3272
• 1 or 2 channels for CT6700 and 3270 series
• ±12 V output
Leak current (2)
AC LEAKAGE CLAMP METER CM4003 (Output and external power supply function)Learn more
0Measures leakage 0.060 mA (resolution 10 μA) to load 200.0 A
0Output function (waveform/RMS), external power supply
0Easy visualization of intermittent trip events (when the Z3210 is installed)
AC ADAPTER Z1013
100 V to 240 V AC