The Hanna HI97106 Chemical Oxygen Demand (COD) Portable Photometer is a rugged, lab-grade instrument designed for field or bench testing of water and wastewater. Here's a complete overview:
Waterproof & durable: Sealed optical system within an IP67-rated floating case—ideal for outdoor conditions
High-precision optics: Dual LEDs (420 /nm for low range; 610 /nm for mid/high/ultra-high) with narrow-band (±1 /nm) interference filters and silicon photocell detector
Consistent readings: Positive-lock vial system ensures exact positioning for repeatability
User guidance: Integrated tutorial walks through sample prep and reagent use
Performance verification: CAL Check™ function with NIST-traceable cuvettes confirms instrument accuracy on-site
Auto data logging: Stores up to 200 measurements, with auto-off after inactivity to save power
Range | Scale (as O ) | Resolution | Accuracy |
---|---|---|---|
LR | 0–150 /mg/L | 1 /mg/L | ±5 /mg/L or ±4 /% at 25 /°C |
MR | 0–1,500 /mg/L | 1 /mg/L | ±15 /mg/L or ±4 /% at 25 /°C |
HR | 0–15,000 /mg/L | 1 /mg/L | ±150 /mg/L or ±2 /% at 25 /°C |
UHR | 0–60.0 /g/L | 0.1 /g/L | ±0.5 /g/L ±3 /% at 25 /°C |