Features:
* Most samples don't need dilute.
* Microscale sample test, minimum 0.5I.
* PC control, more accurate and flexible.
* Ultromicro and cuvette dual platform (Only for BK-CW2000).
* Wide test area, 100 times of the traditional spectrophotometer.
* Test directly without warming up, do not need detection container, lower daily consumption.
Technical Parameters:
| 
			 Model  | 
			
			 BK-CW1000  | 
			
			 BK-CW2000  | 
		
| 
			 Minimum Sample Capacity  | 
			
			 0.5μl  | 
		|
| 
			 Ultromicro Optical Distance  | 
			
			 1mm/0.2mm  | 
			
			 1 mm/0.2mm/0.05mm  | 
		
| 
			 Cuvette Optical Distance  | 
			
			 /  | 
			
			 10mm/5mm/2m m/1 mm  | 
		
| 
			 Whole Wavelength  | 
			
			 19001100nm  | 
		|
| 
			 Scanning Wavelength  | 
			
			 1900850nm  | 
		|
| 
			 Wavelength Accuracy  | 
			
			 1nm  | 
		|
| 
			 Wavelength Resolution  | 
			
			 ≤0.3nm (FWHM @ Hg 253.7nm)  | 
		|
| 
			 Absorbance Accuracy  | 
			
			 2%(0.76 @ 257nm)  | 
		|
| 
			 Absorbance Resolution  | 
			
			 0.002Abs (1mm)  | 
		|
| 
			 Absorbance Range  | 
			
			 0.020300(10mm)  | 
			
			 0.005-300 (10mm)  | 
		
| 
			 Test Accuracy  | 
			
			 0.1-5 Abs.±0.1 Abs 5080 Abs.±2%  | 
			
			 0.0100.5 Abs. ±0.01 Abs 0.5-80 Abs.±2%  | 
		
| 
			 Test Time  | 
			
			 5s  | 
		|
| 
			 Detector  | 
			
			 3648 pixel linear CCD array  | 
		|
| 
			 Light Source  | 
			
			 Xenon flash lamp  | 
		|
| 
			 Sample Test Platform Material  | 
			
			 304 stainless steel and quartz  | 
		|
| 
			 Standard Accessories  | 
			
			 • Driver software CD 1 pc • Power line 1 pc 0 Power Adapter 1 pc  | 
		|
| 
			 Consumption  | 
			
			 24W  | 
		|
| 
			 Power Supply  | 
			
			 AC110/220V±10%, 50/60HZ  | 
		|
| 
			 External Size(W*D*H)mm  | 
			
			 200*130*136  | 
		|
| 
			 Package Size(W*D*H)mm  | 
			
			 310*310*310  | 
		|
| 
			 Net Weight(kg)  | 
			
			 2.5  | 
		|
| 
			 Gross Weight(kg)  | 
			
			 5  | 
		|