The Rocklabs Benchtop RSD (Rotating Sample Divider) is designed to be placed on a work bench and can be used in conjunction with Benchtop Ring Mill.
Big inputs and accurate outputs for seriously reliable results.
Benefits:
| 
			 Specifications 
			 | 
			
			 | 
		
| 
			 Motor power 
			 | 
			
			 0.1kW, single phase 
			 | 
		
| 
			 Maximum input particle size 
			 | 
			
			 10mm 
			 | 
		
| 
			 Width x Height x Depth 
			 | 
			
			 600 x 450.2 x 420mm 
			 | 
		
| 
			 Weight 
			 | 
			
			 58.82kg 
			 | 
		
| 
			 Hopper capacity 
			 | 
			
			 5kg 
			 | 
		
| 
			 Dust extraction 
			 | 
			
			 n/a 
			 | 
		
| 
			 Options 
			 | 
			
			 n/a 
			 | 
		
| 
			 Warranty 
			 | 
			
			 ROCKLABS Bench Top RSD is sold with a 2000-operating-hours or one year warranty (whichever comes first), covering all parts (excluding wear parts) and labour if used under laboratory conditions. 
			 | 
		
| 
			 Comments on taking a representative split 
			 | 
			
			 Taking a representative split from a large sample is not as simple as many people believe. Rolling on a mat, multiple grabs or riffle splitting do not produce accurate results. A classic study* conducted at Bradford University demonstrated that a rotating sample divider provided the best results. 
			* Critical evaluation of powder sample procedures. Allen T and Khan A. A, The chemical Engineer, May 1979, pp108 - 112  | 
		
| 
			 Dividing 
			 | 
			
			 Standard deviation of samples % 
			 | 
		
| 
			 Cone and quartering 
			 | 
			
			 6.810 
			 | 
		
| 
			 Scoop sampling 
			 | 
			
			 5.140 
			 | 
		
| 
			 Riffle splitting 
			 | 
			
			 1.010 
			 | 
		
| 
			 Table sampling 
			 | 
			
			 2.090 
			 | 
		
| 
			 Rotating splitting 
			 | 
			
			 0.125 
			 | 
		
| 
			 Random Variation 
			 | 
			
			 0.076 
			 | 
		
Malaysia