To optimize these specifications for a professional website, I have streamlined the technical terminology and organized the data into a high-impact, easy-to-read format. This layout emphasizes the precision and safety of the YTHG series, which are the primary concerns for lab managers.
The YTHG-072 and YTHG-162 are engineered for rigorous environmental simulation, offering industry-leading stability for electronics, automotive, and pharmaceutical testing. Featuring a robust SUS#304 stainless steel interior and a sophisticated PID control system, these chambers provide a reliable environment for long-term stress testing and quality validation.
| Specification | YTHG-072 | YTHG-162 |
| Internal Volume | 72 Liters | 162 Liters |
| Interior Dimensions (cm) | 45 × 40 × 40 | 60 × 45 × 60 |
| Exterior Dimensions (cm) | 100 × 200 × 170 | 120 × 215 × 185 |
| Interior Material | High-Grade SUS#304 Stainless Steel | High-Grade SUS#304 Stainless Steel |
| Exterior Material | Powder-Coated Cold-Rolled Steel | Powder-Coated Cold-Rolled Steel |
Temperature Range: -70°C to +150°C (Custom ranges available upon request).
Humidity Range: 20% to 98% RH (Ultra-low humidity options of 5% or 10% RH available).
Ramp Rates:
Heating: 3°C/min (Average)
Cooling: 1°C/min (Average)
Precision Control:
Stability/Fluctuation: +-0.3°C | +-0.5% RH
Uniformity: +-1.5°C | +-pm3% RH (Measured at empty load after 30 mins stabilization)
Resolution: 0.01°C | 0.1% RH
Controller: Intuitive Touch Screen Interface with SSR + PID Intelligence.
Airflow: High-efficiency centrifugal fan circulation system for uniform distribution.
Dehumidification: Advanced ADP critical dew point cooling technology.
Refrigeration: Heavy-duty compressor with finned evaporator and air/water-cooled condenser.
Insulation: Dual-layer PU and high-density fiberglass wool for maximum thermal retention.
Compressor: Overload, high/low voltage, and phase sequence protection.
Heating/Humidification: Overheat and water shortage protection with delay logic.
System Safety: Over-temperature cutout, fan overcurrent protection, and visual overload indicators.
Vacuum-sealed glass observation window.
(50mm) Access/Test hole.
Integrated interior lighting and status indicators.
Two (2) freely adjustable SUS#304 compartment shelves.
We specialize in non-standard requirements. Whether you need specific dimensions, extended temperature ranges, or specialized port configurations, our engineering team can customize the YTHG series to fit your exact testing protocol.
More detail about OS Electronics Sdn Bhd
Malaysia