| Small Sample Adapter Viscosity Ranges cP (mPa·s) | |||||||||||||
| MODEL | Spindle: SC4-18 Sample Chamber:* SC4-13R(P) Sample Volume: 6.7mL Shear Rate (sec-1): 1.32N |
Spindle: SC4-31 Sample Chamber:* SC4-13R(P) Sample Volume: 9.0mL Shear Rate (sec-1): .34N |
Spindle: SC4-34 Sample Chamber:* SC4-13R(P) Sample Volume: 9.4mL Shear Rate (sec-1): .28N |
Spindle: SC4-16 Sample Chamber:* SC4-8R(P) Sample Volume: 4.2mL Shear Rate (sec-1): .29N |
Spindle: SC4-25Z(316 s/s only) Sample Chamber:* SC4-13R(P) Sample Volume: 16.1mL Shear Rate (sec-1): .22N |
||||||||
| DV3TLV*** | 1.2-30K | 12-300K | 24-600K | 48-1.2M | 192-4.8M | ||||||||
| DV2TLV**** | 1.5-30K | 15-300K | 30-600K | 60-1.2M | 240-4.8M | ||||||||
| DV1LV | 3-10K | 30-100K | 60-200K | 120-400K | 800-1.6M | ||||||||
| DVELV | 3-10K | 30-100K | 60-200K | 120-400K | 800-1.6M | ||||||||
| LVT | 5-10K | 50-100K | 100-200K | 200-400K | 800-1.6M | ||||||||
| DV3TRV*** | It is possible | ||||||||||||
| DV2TRV**** | to use the above spindles with any of these instruments. | ||||||||||||
| DV1RV | Digital Viscometers/Rheometers will automatically calculate | ||||||||||||
| RVDVE | viscosity. Please contact Brookfield or an authorized dealer | ||||||||||||
| RVT | if you require information on viscosity ranges. | ||||||||||||
| DV3THA*** | NA | ||||||||||||
| DV2THA**** | NA | ||||||||||||
| DV1HA | NA | ||||||||||||
| HADVE | NA | ||||||||||||
| HAT | NA | ||||||||||||
| DV3THB*** | NA | ||||||||||||
| DV2THB**** | NA | ||||||||||||
| DV1HB | NA | ||||||||||||
| HBDVE | NA | ||||||||||||
| HBT | NA | ||||||||||||
| Small Sample Adapter Viscosity Ranges cP (mPa·s) | ||||||||||||||
| MODEL | Spindle: SC4-21 Sample Chamber:* SC4-13R(P) Sample Volume: 7.1mL Shear Rate (sec-1): .93N |
Spindle: SC4-27 Sample Chamber:* SC4-13R(P) Sample Volume: 10.4mL Shear Rate (sec-1): .34N |
Spindle: SC4-15 Sample Chamber:* SC4-7R(P) Sample Volume: 3.8mL Shear Rate (sec-1): .48N |
Spindle: SC4-28 Sample Chamber:* SC4-13R(P) Sample Volume: 11.0mL Shear Rate (sec-1): .28N |
Spindle: SC4-29 Sample Chamber:* SC4-13R(P) Sample Volume: 13.5mL Shear Rate (sec-1): .25N |
Spindle: SC4-14 Sample Chamber:* SC4-6R(P) Sample Volume: 2.1mL Shear Rate (sec-1): .40N |
||||||||
| DV3TLV*** | Not applicable for historical reasons. However, it is possible | |||||||||||||
| DV2TLV**** | to use the above spindles with any of these instruments. | |||||||||||||
| DV1MLV | Digital Viscometers/Rheometers will automatically calculate | |||||||||||||
| LVDVE | viscosity. Please contact Brookfield or an authorized dealer | |||||||||||||
| LVT | if you require information on viscosity ranges. | |||||||||||||
| DV3TRV*** | 20-500K | 100-2.5M | 200-5M | 200-5M | 400-10M | 500-12.5M | ||||||||
| DV2TRV**** | 25-500K | 125-2.5M | 250-5M | 250-5M | 500-10M | 625-12.5M | ||||||||
| DV1RV | 50-170K | 250-830K | 500-1.7M | 500-1.7M | 1K-3.3M | 1.25K-4.2M | ||||||||
| RVDVE | 50-170K | 250-830K | 500-1.7M | 500-1.7M | 1K-3.3M | 1.25K-4.2M | ||||||||
| RVT | 50-100K | 250-500K | 500-1M | 500-1M | 1K-2M | 1.25K-2.5M | ||||||||
| DV3THA*** | 40-1M | 200-5M | 400-10M | 400-10M | 800-20M | 1K-25M | ||||||||
| DV2THA**** | 50-1M | 250-5M | 500-10M | 500-10M | 1K-20M | 1.25K-25M | ||||||||
| DV1HA | 100-300K | 500-1.7M | 1K-3.3M | 1K-3.3M | 2K-6.7M | 2.5K-8.3M | ||||||||
| HADVE | 100-300K | 500-1.7M | 1K-3.3M | 1K-3.3M | 2K-6.7M | 2.5K-8.3M | ||||||||
| HAT | 100-200K | 500-1M | 1K-2M | 1K-2M | 2K-4M | 2.5K-5M | ||||||||
| DV3THB*** | 160-4M | 800-20M | 1.6K-40M | 1.6K-40M | 3.2K-80M | 4K-100M | ||||||||
| DV2THB**** | 200-4M | 1K-20M | 2K-40M | 2K-40M | 4K-80M | 5K-100M | ||||||||
| DV1HB | 400-1.3M | 2K-6.7M | 4K-13.3M | 4K-13.3M | 8K-26.7M | 10K-33.3M | ||||||||
| HBDVE | 400-1.3M | 2K-6.7M | 4K-13.3M | 4K-13.3M | 8K-26.7M | 10K-33.3M | ||||||||
| HBT | 400-800K | 2K-4M | 4K-8M | 4K-8M | 8K-16M | 10K-20M | ||||||||
N/A=Not applicable for historical reasons. However, it is possible to use any spindle/chamber combination with any torque ranges. Digital viscometers/rheometers will automatically calculate viscosity.
** Disposable chamber available in 13R size and requires SC4-45YD water jacket
*** Ranges also apply to previous “DV-III” series Rheometers
****Ranges also apply to previous “DV-II” series Rheometers
Examples
SC4-13R Sample Chamber
SC4-13RPY Sample Chamber with RTD temperature probe and cable to viscometer/rheometer
SC4-27 Stainless Steel Spindle
SC4-13RP Sample Chamber with RTD temperature probe
SC4-13RD-100 Disposable Sample Chamber available in packages of 100
SC4-27D Disposable Spindle
Note: Solid shaft spindles available: SC4-27SD, SC4-28SD, SC4-29SD. Hastelloy C available for some spindles/chambers-call for details.