Technical performance
1. Working conditions
L environment temperature: 5 ~ 35 ℃ 邋 邋 relative humidity: < 85%
L power supply: 220V± 10% of 50 hz & plusmn; 1 hz
L has no significant vibration
L no other strong magnetic field interference except the magnetic field.
2. Main technical indicators
L measurement range: conductivity: (0 ~ 199.9)μ S/cm; (200 ~ 1999) & mu; S/cm;
(2.00 ~ 19.99) ms/cm; (20.0 ~ 199.9) ms/cm;
L resolution: 0.1/1μ S/cm; 0.01/0.1 ms/cm;
L accuracy: electrical meter :± 1.0% FS. Matching: & plusmn; 1.5% FS
L temperature compensation scope: 0 ~ 50 ℃ automatic temperature compensation
L temperature number: 2.0% / ℃
Specification constants: 0.1, 1 and 10
L data storage: 50 groups
Storage: number, measurement value, measurement unit and temperature value
L power supply: DC9V adapter
L size and weight: 220× 200 & times; 80 mm (long & times; Wide & times; High) 0.7 kg
More detail about MTM Precision Sdn Bhd