| Measurement Characteristics | |
|---|---|
| Principal of Operation |
Redox Electrode with Extended Reference Path Temperature: Silicon Based Linear Thermistor |
| Applicable Units | mV, °C |
| Range |
Redox: -1999 to +1999 mV Temperature: 0 to 50°C |
| Accuracy |
Redox: +/- 2 mV Temperature: +/- 0.5°C |
| Resolution |
Redox: 1 mV Temperature: 0.1°C |
| Service Requirement | |
| Cleaning | See FlexTech range of autoclean mounting options |
| Sensor Service | Electrode Replacement - typically yearly |
|
Physical |
|
|---|---|
| Dimensions | 270 x Ø60 mm |
| Weight | 1.2 kg inc 10 metres of cable |
| Wetted Materials | Body: Black PVC, Nylon Cable Gland, Electrode: PPS (Ryton) |
| Seal Material | Nitrile |
| Enclosure Rating | IP68 |
| Cable Type | 4 core, 2 twisted pair, 5 mm O/D Polyurethane Coated |
| Cable Length | 10 metres standard, 100 metres maximum |
| Temperature Rating | 0 to +50°C |
| Pressure Rating | 2 Bar |
| Depth Rating | 20 metres |
| Thread for Mounting Shaft Connection | 1 1/2'' BSPP Male |
| Interface To Monitor | |
| Power Supply | 12VDC from Monitor |
| Type | ModTech - RS485 |
| Physical | |
|---|---|
| Dimensions (hxwxd) | 235 x 310 x 155 mm |
| Weight | 0.35 kg |
| Protection Class | IP68 |
| Enclosure Material | Black Acetal Co-Polmer with Stainless Steel Mounting PLate |
| Cable Entry | Integral Cable Gland |
| Cable Length | 10 metres standard, 100 metres maximum |
| Interface to Monitor | |
| Power Supply | 12 VDC from Monitor |
| Sensor Output | ModTechw², RS485 |
| Environmental Data | |
| Operating/Storage Temperature | 0 to 70ºC |
| Location | Indoor/Outdoor |
| Maintenance | |
| Service |
Electrode Replacement Periodic cleaning is required at a frequency determined by the application. Our self cleaning shaft option is highly recommended, this eliminates manual cleaning between calibrations. |
| Calibration | The sensor is inherently stable, calibration should be carried out to suit the application. Our recommendation is every 6 months. |