Thermocouple-Resistance Thermometer
| 416M | - | □ | - | □ | - | □ | - | □ |
| ① | ② | ③ | ④ |
| Code | Type *1 | Measuring Range | Display Range | Accuracy *2 |
|---|---|---|---|---|
| TC | R | 100 to 1700°C | -50 to 1750°C | ±(0.3% of rdg + 1°C) |
| K | -100 to 1300°C | -199 to 1350°C | ||
| E | -130 to 1000°C | -199 to 1050°C | ||
| J | -140 to 1200°C | -199 to 1250°C | ||
| T | -199 to 400°C | -199 to 420°C | ||
| B | 600 to 1800°C | -20 to 1810°C | ||
| N | -100 to 1300°C | -199 to 1350°C |
| *1 There will be °C display of K type TC during shipment. | |
| *2 Accuracy | : Defined in the measuring range, at 23 ± 5°C, 45 to 75% RH. |
| Temperature coefficient | : ±300 ppm/°C (operating temp.: 0 to 50°C) |
| Cold junction compensation | : ±1°C (operating temp.: 0 to 50°C) |
| Calibration is made with standard thermo-electricity input (mV) in accordance with JIS C-1602. | |
| Code | Type *3 | Measuring Range | Display Range | Accuracy *4 |
|---|---|---|---|---|
| PT | Pt100Ω | -199.9 to 600.0°C | -199.9 to 650.0°C | ±(0.2% ofrdg + 0.3°C) |
| JPt100Ω |
| *3 There will be Pt100Ω and °C setting display during shipment. | |
| *4 Accuracy | : Defined in the measuring range, at 23 ± 5°C, 45 to 75% RH. |
| Temperature coefficient | : ±200 ppm/°C (operating temp.: 0 to 50°C) |
| Calibration is made with standard resistance values in accordance with JIS C-1604. | |
| Code | Specifications | Output Impedance | Allowable Load Resistance |
|---|---|---|---|
| Blank | For display only | ||
| 03 | Analog output DC 0 to 1 V | Less than 0.1 Ω | More than 1 kΩ |
| 04 | Analog output DC 0 to 5 V | Less than 0.1 Ω | More than 5 kΩ |
| 05 | Analog output DC 0 to 10 V | Less than 0.1 Ω | More than 10 kΩ |
| 09 | Analog output DC 1 to 5 V | Less than 0.1 Ω | More than 5 kΩ |
| 23 | Analog output DC 0 to 1 mA | More than 5 MΩ | 0 to 5 kΩ |
| 29 | Analog output DC 4 to 20 mA | More than 5 MΩ | 0 to 600 Ω |
| Display | Red or green LED (character height 15mm) with zero-suppress function. |
|---|---|
| Mode display | 1 digit, red or green LED (character height 8mm) |
| Over-range display | When exceeded the displayable range, the minimum or maximum value of the displayable range flashes. |
| Open circuit of input |
Thermocouple input: Flashes with the minimum value of the display range. Resistance thermo bulb input: Flashes with the maximum value of the display range. |
| Resolution |
Thermocouple: 1°C Resistance thermometer: 0.1°C |
| External resistance |
Thermocouple: Less than 500 Ω Resistance thermometer: Less than 5 Ω/lead wire. |
| Overload | DC ±10 V |
| Sampling rate | Approx. 2.5 times/s |
| Input configuration | Single ended, floating input |
| A/D conversion | ⊿-∑ conversion system |
| Noise rejection |
Normal mode (NMR): More than 50 dB Common mode (CMR): More than 110 dB Power source line mixing noise: 1000 V |
| Hold function |
Holds measured data, peak/bottom memory value and data output (option). (Not isolated from input) |
| Peak/bottom memory function | Display of maximum or minimum value is selectable with the front panel switch |
| Withstand voltage | |
| Insulation resistance | DC 500 V More than 100 MΩ |
| Operating temperature | 0 to 50°C |
| Protection degree | IP65 |
| Weight | Approx. 200 g |