| Item Code | 200000 |
|---|---|
| Model | DP810 |
| Internal dimensions(W×D×H mm) | 800×800×800mm |
| External dimensions(W×D×H mm) | 1020×1019×1850mm |
| Remarks(dimensions) | 512L |
| Power Source |
AC220VV 30A Single-phase |
| Weight | 600kg |
| Method | Vacuum drying by decompressed chamber direct heating |
| Operating temperature range | 40 to 200℃ |
| Operating pressure range | 100 to 0.1kPa(760 to 1 Torr) |
| Temp. adjustment accuracy | ±1℃(at200℃) |
| Interior Material | Stainless steel |
| Exterior Material | Cold rolled steel plate with baked-on melamine resin finish |
| Door | Single swing door |
| Heat insulating material | Rock wool |
| Heater | Mica heater, 6.5kW |
| Operation functions | Fixed temp., Auto-start, Auto-stop, Quick auto stop, Program (max.99 steps, 99 patterns, repeat) |
| Heater circuit control | Triac zero-cross control |
| Temp. sensor | K-thermocouple |
| Exhaust port | NW40 |
| Inlet/Purge port | RC 1/4 |
| Temp. control method | PlD control by microprocessor |
| Temp. setting method | Digital setting with UP/DOWN key |
| Temp display method | Digital display |
| Other display | Temperature indicating the operation state displayed by LED |
| Additional functions | Calendar timer (max. 24 Hrs.), Integration time (max. 65535 hrs.), Clock, Calibration off-set, Display the amount of power consumption / CO2 discharge / Heater operation amount,Power failure recovery mode, User setting information save and recall |
| Safety countermeasures | Self diagnosis functions(Sensor, Heater, Triac, Automatic overheating prevention), Independent overheating prevention,Key lock function, Electric leakage breaker |
| Accessories Shelf plate | Punched stainless steel, 2 pcs |