| Item Code | 212862 |
|---|---|
| Model | DVS402 |
| Method | Natural gravity convection |
| Operating temp. range | Room temp.+5~260℃ |
| Temp. control accuracy | ±1℃(at 260℃) |
| Temp. distribution accuracy | ±5℃(at 260℃) |
| Max. temp. reaching time | Approx. 90 min. (Room temp. +5℃ to 260℃) |
| Interior material | Stainless steel |
| Exterior material | Cold rolled steel plate with melamine resin baking finish |
| Heat insulating material | Glass wool |
| Heater | Stainless pipe heater |
| Heater | 1.2kW |
| Observation window | 250×280 mm Chemically strengthened glass×3 |
| Cable hole | 30 mm l.D.×1 pcs.(right side) |
| Exhaust port | 30 mm l.D.×2 pcs.(on top) |
| Temp. controller | 3 patterns program controller, PID control by microprocessor |
| Temp. setting method | Operation menu key and digital setting by UP/DOWN key |
| Temp display method | Measurement temp. : Digital display by green LED Setting temp. : Digital display by red LED |
| Timer | 1 min. to 99 hrs. 59 min. and 100 hrs. to 999 hrs. 50 min. |
| Operation function | Fixed temperature operation, Program operation, Auto start, Auto stop, Quick Auto-stop |
| Program mode | Program operation: 6 pattern, 30 steps (30 steps×1, 15 steps×2, 10 steps×3) |
| Additional functions | Calibration off-set function, Key lock, Uninterruptible power for memory, Pattern repeat function |
| Heater circuit control | SSR control |
| Sensor | K-thermocouple |
| Safety device | Self diagnosis functions (Temp. sensor abnormal, Heater disconnection, SSR- short, Memory abnormal |
| Safety device | Automatic overheating prevention), Overheat prevention, Electric leakage breaker with over current protection |
| Shelf plate load | Approx. 15kg / pcs. |
| Shelf support step number / pitch | 9 steps / 30mm |
| Shelf plate / Shelf bracket | Stainless steel, 2 pcs. / 4 pcs. |
| Internal dimensions(W×D×H mm) | 450×490×450mm |
| External dimensions(W×D×H mm) | 560×601×820mm |
| Remarks(dimensions) | 99L |
| Power Source |
AC220V 6A Single-phase |
| Weight | 48kg |