| Item Code | 212158 |
|---|---|
| Model | DP410 |
| Method | Vacuum drying by decompressed chamber direct heating |
| Max. temp. reaching time | Approx. 80 min. |
| Interior material | Stainless steel |
| Exterior material | Cold rolled steel plate with baked-on melamine resin finish |
| Door | Single swing door |
| Heat insulating material | Rock wool |
| Heater | Mica heater, 2.25 kW |
| Vacuum gauge | Bourdon tube type, 0 to 0.1 MPa (Gauge pressure) |
| Observation window | Tempered glass and polycarbonate resin plate |
| Temp. control method | PID control by microprocessor |
| Temp. setting method | Digital setting with UP/DOWN key |
| Timer | 0 min. to 99 Hrs. 59 min. and 100 Hrs. to 999 Hrs. |
| Min. division | 1 min. or 1 Hrs. |
| Operation function | Fixed temperature operation, Auto-start operation, Auto-stop operation |
| Additional functions | Calendar timer (max. 24 Hrs.), Integration time (max 65535 Hrs.), Time display |
| Heater circuit control | Triac zero-cross control |
| Temp. sensor | K-thermocouple (Double sensor) |
| Exhaust port / Purge port | NW25 flange / Rc 1/4 (18mm O.D.) |
| Operating temperature range | +40 to 200°C |
| Operatingpressure range | 101 to 0.1 kPa (760~1 Torr) at absolute pressure |
| Temp. Fluctuation | ± 1.5°C (at 200°C) |
| Pump connection port / Purge port | NW25 flange / Rc 1/4 |
| Temp display method | Digital display by green LED |
| Safety countermeasures | Self diagnosis functions (Sensor, Heater, Triac, Automatic overheating prevention), Independent overheating prevention, Key lock function, Electric leakage breaker |
| Leakage breaker | 20A |
| Independent cut-out circuit | Set temperature range: 0 to 230°C |
| Shelf support step number | 4 steps |
| Shelf support pitch | 105mm |
| Accessories Shelf plate | Punched stainless steel, 2 pcs |
| Internal dimensions(W×D×H mm) | 450×450×450mm |
| External dimensions(W×D×H mm) | 670×669×1500mm |
| Remarks(dimensions) | 91L |
| Power Source |
AC220V 11A Single-phase |
| Weight | 210kg |
Malaysia