| range | resolution | Accuracy | remarks | |
|---|---|---|---|---|
| ACV | 60 mV | 0.01 mV | ±(1.0% + 3 dgt) |
|
| 600 mV | 0.1 mV | |||
| 6V | 0.001 V | |||
| 60 V | 0.01 V | |||
| 600 V | 0.1 V | |||
| 750 V | 1 V | |||
| DCV | 60 mV | 0.01 mV | ±(0.5% + 3 dgt) |
|
| 600 mV | 0.1 mV | |||
| 6V | 0.001 V | |||
| 60 V | 0.01 V | |||
| 600 V | 0.1 V | |||
| 1000 V | 1 V | |||
| ACA | 60mA | 0.01mA | ±(1.5% + 3 dgt) |
|
| 600mA | 0.1mA | |||
| 6A | 0.001 A | |||
| 10A | 0.01 A | |||
| DCA | 60mA | 0.01mA | ±(1.2%rdg + 3dgt) |
|
| 600mA | 0.1mA | |||
| 6A | 0.001 A | |||
| 20A | 0.01 A | |||
| Ω | 600 ohms | 0.1 Ω | ±(0.5%rdg + 3dgt) |
|
| 6 kΩ | 0.001 kΩ | |||
| 60 kΩ | 0.01 kΩ | |||
| 600 kΩ | 0.1 kΩ | |||
| 6 MΩ | 0.001 MΩ | |||
| 60 MΩ | 0.01 MΩ | ±(1.5%rdg + 3dgt) | ||
| capacitance | 9.999 nF | 0.001 nF | ±(5.0%rdg + 20dgt) |
|
| 99.99 nF | 0.01 nF | ±(2.0%rdg + 5dgt) | ||
| 999.9 nF | 0.1 nF | |||
| 9.999 μF | 0.001 μF | |||
| 99.99 μF | 0.01 μF | |||
| 999.9 μF | 0.1 μF | |||
| 9.999 mF | 0.001 mF | ±(5.0%rdg + 5dgt) |
| Buzzer Threshold | remarks | |
|---|---|---|
| Continuity check | Approximately 50 Ω |
|
| range | resolution | remarks | |
|---|---|---|---|
| Diode Check | 0 V to 3 V | 1 mV |
|
| Measurement range | Accuracy | frequency | |
|---|---|---|---|
| frequency | 1 Hz to 9.999 MHz | ±(0.1%rdg + 2dgt) | 1.0 Vrms or more, sine wave, 50% square wave |
| DUTY | 5.0 to 99.0% | ±(0.1% + 2 dgt) | 10 Hz to 1 kHz, 1.0 Vrms or more, square wave |
| Detection method | True RMS |
|---|---|
| display | LCD, maximum display “6000” |
| Range switching | Auto |
| Overrange display | "OL" mark display |
| Low battery indicator | When the battery voltage drops, a battery symbol is displayed. |
| polarity display | Automatic switching, only "-" displayed |
| sampling | Approximately 3 times/second |
| Auto Power Off | Approximately 15 minutes (can be canceled) |
| Operating temperature and humidity | 0 to +40°C, 75% RH or less (no condensation) |
| Storage temperature and humidity | -20 to +60°C, 80% RH or less (no condensation) |
| size | Approx. W 65 × H 130 × D 32 mm |
| weight | Approximately 130 g (including batteries) |
| Test lead length | Approximately 75cm |
| power supply | 2 x AAA batteries |
| battery life | Approximately 400 hours (when using alkaline batteries) |
| accessories | Instruction manual, test leads |
| option |
|
| proofreading | Possible |
| Compliance standards | IEC61010-1 CAT II 1000V, CAT III 600V |
| JAN | 4983621100471 |
Malaysia