| Model | TMT-900S (Water cooled) |
TMT-125F (Air cooled) |
TMT-216F (Air cooled) |
TMT-512F (Air cooled) |
| Internal size (D*W*H)mm | 950*950*850 | 500*500*500 | 600*600*600 | 800*800*800 |
| External size (D*W*H)mm | 1750*1500*2100 | 990*1130*1640 | 1200*1230*1800 | 1400*1430*2000 |
| Temperature Range | Ambient temperature +10°C ~ 80°C | |||
| Humidity Range | 45 ~ 95%RH | |||
| Black Panel Temperature | 63°C ~ 70°C | |||
| Temperature Fluctuation | ±0.5°C | |||
| Temperature Uniformity | ±2.0°C | |||
| Rate of Rice of Temperature | RT to 100°C within 30 mins | |||
| Humidity Uniformity | ±1.0%RH | |||
| Rainfall Time | 0 ~ 999min, continuous rainfall is adjustable | |||
| Rainfall Cycle | 0 ~ 240min, interval rainfall is adjustable | |||
| Irradiance | 550w/m² | |||
| Heater Power | 2KW | |||
| Humidifier Power | 2KW | |||
| Distance from specimen shelf and xenon lamp | 300 ~ 375mm | |||
| Specimen Shelf | Round and rotatable, rotation speed is adjustable | |||
| Xeron Lamp | Wavelength 290~800nm, and allocate borosilicate glass filter to making the lamp similiar to the sunlight | |||
ISO 11341《Paints and varnishes-Artificial weathering and exposure to artificial radiation–Exposure to filtered xenon-arc radiation》
ISO 12040 《Graphic Technology – Prints and Printing Inks – Assessment of Light Fastness Using Filtered Xenon Arc Light 》
ISO 16474-1 《Paints and varnishes — Methods of exposure to laboratory light sources — Part 1: General guidance》
ISO 16474-2 《Paints and varnishes — Methods of exposure to laboratory light sources — Part 2: Xenon-arc lamps》
ASTM D3451 《Standard Guide for Testing Coating Powders and Powder Coatings》
ASTM D3794《Standard Guide for Testing Coil Coatings》
ASTM D4303 《 Standard Test Methods for Lightfastness of Pigments Used in Artists’ Paints》
ASTM D5010 《Standard Guide for Testing Printing Inks and Related Materials》
ASTM D6577 《Standard Guide for Testing Industrial Protective Coatings》
ASTM D6695 《Standard Practice for Xenon-Arc Exposures of Paint and Related Coatings》
ASTM G151《Standard Practice for Exposing Nonmetallic Materials in Accelerated Test Devices that Use Laboratory Light Sources》
ASTM G155《Practice for Operating Xenon Arc Light Apparatus for Exposure of Non-Metallic Materials》
ISO 4892-1 《Plastics – Methods of Exposure to Laboratory Light Sources – Part 1: General guidance》
ISO 4892-2《Plastics – Methods of Exposure to Laboratory Light Sources – Part 2: Xenon-arc lamps》
SAE J2412 《Accelerated Exposure of Automotive Interior Trim Components Using a Controlled Irradiance Xenon-Arc Apparatus》
We deliver custom-made chambers
We will meet your demands to design chamber with good performance according to customer standared and other special requirement