Features:
*7-inch LCD display indicates temperature, gas flow and heating curve.
*SelMest linction and automatic fault recognition function.
*8 pcs external extension interfaces can connect with control valve.
*Equipped with a flame ionization detector (FID).
*Multi-core and 32-bit embedded hardware system ensures the reliable operation.
*Synchronous external trigger function and the external signal (auto sampler, thermal analyzer, etc.) can start the equipment and workstation software at the same time.
*With chromatography workstation software to process data.
*Memory function: can store 20 sample test modes.
*One key start up function.
 
Technical Parameters:
| 
			 Model  | 
			BK-GC112A | |
| 
			 Oven  | 
			
			 Capacity  | 
			
			 22L  | 
		
| 
			 Temp. Range  | 
			
			 RT+5°C~400°C  | 
		|
| 
			 Temp. Accuracy  | 
			
			 ±0.1 °C  | 
		|
| 
			 Temp. Program  | 
			
			 9 step  | 
		|
| 
			 Program Total Time  | 
			
			 9999.9 min  | 
		|
| 
			 Max Temp. Heating Rate  | 
			
			 60°C/min  | 
		|
| 
			 Sample Injector  | 
			
			 Temp. Range  | 
			
			 RT+7°C~420°C  | 
		
| 
			 Column or split injector can be installed  | 
		||
| 
			 Carrier gas flow control type: Constant pressure mode  | 
		||
| 
			 Max. Quantity  | 
			
			 3pcs  | 
		|
| 
			 Pre Column Pressure Range  | 
			
			 0-400kpa  | 
		|
| 
			 Pre Column Pressure Accuracy  | 
			
			 0.1 kpa  | 
		|
| 
			 Flow Setting Range  | 
			
			 H2: 0-200ml/min  | 
		|
| 
			 N2: 0-150ml/min  | 
		||
| 
			 Detector  | 
			
			 Temp. Range  | 
			
			 RT+7°C~420°C  | 
		
| 
			 Hydrogen Flame Ionization Detector (FID)  | 
			
			 Logarithmic amplifier  | 
		|
| 
			 High voltage switch control  | 
		||
| 
			 Baseline signal display  | 
		||
| 
			 Detection limit: ^3*1012g/s (Sample: n-hexadecane)  | 
		||
| 
			 Baseline noise: <5*1014A  | 
		||
| 
			 Baseline draft:《6*1 O ^A  | 
		||
| 
			 Dynamic range: 107  | 
		||
| 
			 RSD:£3%  | 
		||
| 
			 Thermal Conductivity Detector (TCD)(Optional)  | 
			
			 Sensitivity: 8000mv.ml/mg(Sample: n-hexadecane)  | 
		|
| 
			 Baseline noise: <0.05mV  | 
		||
| 
			 Baseline draft: <0.15mV/30min  | 
		||
| 
			 Dynamic range: 105  | 
		||
| 
			 Power Supply  | 
			
			 AC220V±10%, 50Hz  | 
		|
| 
			 Standard Accessory  | 
			
			 FID detector, Capillary injection system, Chromatography workstation software,RS232 port, LAN port  | 
		|
| 
			 Optional Accessory  | 
			
			 TCD detector, Chromatographic column, Gas purifier  | 
		|
| 
			 Package Size(W*D*H)  | 
			
			 735*725*835mm  | 
		|
| 
			 Gross Weight  | 
			
			 80kg  | 
		|