| eature | Specification |
|---|---|
| Output | 0.0 to 100.0 kV DC (100V resolution) 1.0mA Max (continuous, at max V) |
| Output Voltage Accuracy | 1%, <0.5% ripple (20-100kv) |
| Current Measurement | 0-1100uA, +/-1uA resolution, 1% accuracy at FS |
| Test Duration (user setting) | 1s resolution up to 24 hours, or Continuous |
| Discharge Time | <3 seconds |
| HV Present Warning | High-Brightness Visual Indicator and Audible Alert |
| Pass/Fail Limit (user setting, VI Test mode) | 100μA, 200μA and 300μA |
| Test Stop Limit (user setting, HiPot mode) | 10-1000µA in 10µA steps |
| Input Power | 100-240Vac, 47-63 Hz, 2A |
| Battery Power | HIPO-100B only: 14.4V/3400mAh, Li-ion, user replaceable |
| Display | 4.3” 480x272 IPS TFT, Ultra-bright (visible under bright sunlight) |
| Control | Touch control (gloved hand compatible) + Power/Home/Test switch Emergency Stop button |
| Start-up time | Instrument is ready to use within 2s of power button pressed |
| Test Results Storage | 200 tests |
| Test Results Transfer | Via USB flash drive |
| Remote Control | Via USB virtual COM port (ASCII protocol, using any serial-capable scripting language or test automation package) |
| Date/Time | Internal real-time-clock, battery backup (5 year lifetime, user replaceable) |
| Firmware Updates | Field updatable via USB flash drive |
| Printer | Built-in 2.25-inch Thermal Printer |
| Dimensions | 410x310x150mm |
| Weight | 9.6kg |
| Operating Temperature | 0°C to +50°C |
| Storage Temperature | -30 °C to +70 °C |
| Humidity | 5% — 95% RH, non-condensing |
| Protection Class | IP67 (within transport/operation case, lid closed) |
HIPO-100 Specifications
HVX Products Overview