Colorimetry – LISA enables reliable low-cost color measurements. LISA color uses two different LEDs for long-term stable measurements of SAC or colors at different wavelengths. The second channel is used for turbidity/background correction. The cutting-edge device platform, used in all other TriOS photometers, enables optical path lengths of 50, 100, 150 and 250 mm, so that almost any application can be easily implemented.
| Light Source | 2 LEDs | |
| Detector | Photo diode | |
| Measurement principle | Attenuation, Transmission | |
| Optical path | 50 mm, 100 mm, 150 mm, 250 mm | |
|  | ||
| Parameter | SAC436 or Colouring (based on DIN EN ISO 7887 (410 nm, 436 nm, 525 nm or 620 nm)) or Pt-Co color number (APHA/Hazen) (390 nm or 455 nm) or Cr-Co color number (390 nm or 413 nm) | |
| Measurement accuracy | 0.5 % | |
| Turbidity compensation | at 740 nm | |
| Data logger | ~ 2 MB | |
| T100 response time | min. 4 s | |
| Measurement interval | min. 2 s | |
| Housing material | Stainless steel (1.4571/1.4404) or titanium (3.7035) | |
| Dimensions (L x Ø) | 340 mm x 48 mm (with 50 mm path) | ~ 13.4” x 1.9” (with 50 mm path) | 
| Weight stainless steel | ~ 2.4 kg (with 50 mm path) | ~ 5.3 lbs (with 50 mm path) | 
| Weight titanium | ~ 1.3 kg (with 50 mm path) | ~ 2.9 lbs (with 50 mm path) | 
| Interface digital | Ethernet (TCP/IP), RS-232 or RS-485 (Modbus RTU) | |
| Interface analog | Ethernet (TCP/IP), 4...20 mA | |
| Power consumption | ≤ 1 W | |
| Power supply | 12...24 VDC (± 10 %) | |
| Maintenance effort | ≤ 0.5 h/month (typical) | |
| Calibration/maintenance interval | 24 months | |
| System compatibility | Modbus RTU or Analog Out (4...20 mA) | |
| Warranty | 1 year (EU: 2 years) | US: 2 years | 
| Max. pressure with SubConn | 30 bar | ~ 435 psig | 
| Max. pressure with fixed | 3 bar | ~ 43.5 psig | 
| Max. pressure in FlowCell | 1 bar, 2...4 L/min | ~ 14.5 psig, 0.5 to 1.0 gpm | 
| Protection type | IP68 | NEMA 6P | 
| Probentemperatur | +2...+40 °C | ~ +36 贔 to +104 贔 | 
| Umgebungstemperatur | +2...+40 °C | ~ +36 贔 to +104 贔 | 
| Lagertemperatur | -20...+80 °C | ~ -4 贔 to +176 贔 | 
| Anströmgeschwindigkeit | 0,1...10 m/s | ~ 0.33 fps to 33 fps | 
| Parameter Variations | According to the standard | Unit | Factor | Path Length [mm] | ||||
|  |  |  | 10 | 50 | 100 | 150 | 250 | |
| SAC 436 nm | DIN EN ISO 7887:2011_method B | 1/m | - | 0.5...150 | 0.1...30 | 0.05...15 | 0.03...10 | 0.02...6 | 
| SAC 525 nm | DIN EN ISO 7887:2011_method B | 1/m | - | 0.5...150 | 0.1...30 | 0.05...15 | 0.03...10 | 0.02...6 | 
| SAC 620 nm | DIN EN ISO 7887:2011_method B | 1/m | - | 0.5...150 | 0.1...30 | 0.05...15 | 0.03...10 | 0.02...6 | 
| True Color 410 nm | DIN EN ISO 7887:2011_method C | mg/L Pt | 18.52 | 10...2800 | 2...560 | 1.0...280 | 0.6...185 | 0.4...110 | 
| Pt-Co-Color 390 | DIN EN ISO 6271-2016:05 | mg/L Pt | 7.4 | 4...1100 | 0.8...220 | 0.4...110 | 0.3...75 | 0.2...45 | 
| Pt-Co-Color 455 | DIN EN ISO 6271-2016:05 | mg/L Pt | 36.4 | 20...5500 | 4...1100 | 2.0...550 | 1.5...360 | 0.8...220 | 
| Cr-Co-Color 380 | - | ° (color degree) | 9.7 | 5...1500 | 1...300 | 0.5...150 | 0.3...100 | 0.2...60 | 
| Cr-Co-Color 413 | GOST 3351:1974 | ° (color degree) | 34.1 | 20...5500 | 4...1100 | 2.0...550 | 1.5...360 | 0.8...220 | 
| Turbidity 740 nm | - | FAU** | 6.0 / 0.01242 | 3...330 | 0.6...60 | 0.3...30 | 0.2...20 | 0.12...12 | 
|  |  |  |  |  |  |  | ||
| under laboratory conditions **Formazin Attenuation Unit Version  |  |  |  | |||||