| Application range | up to 405 °C (°C/°F selectable) |
| Ignition | Gas and electric (interchangeable) |
| Heating | According to method or user-defined |
| Stirring | According to method or user-defined |
| Cooling | Built-in forced air (fan) |
| Flash point detection | Thermocouple |
| Sample temperature | Pt100 glass probe |
| Sample temperature calibration |
Dynamic calibration against certified ASTM thermometer and offset capability with correction table (21 calibration points) |
| Barometric pressure correction | Automatic correction |
| Safety | Overheat protection, automatic shut-off, fire-extinguishing system, password |
| Interfaces | 3x USB, 1x RS232, 1x LAN |
| Gas supply |
Test flame: 50 mbar of propane or butane Fire extinguisher: 6 bar to 12 bar of nitrogen or carbon dioxide |
| Power supply | 115 V/230 V, 50 Hz/60 Hz, 1000 W |
| Dimensions | 230 mm x 410 mm x 460 mm (W x D x H) |
| Weight | 14 kg |