| Differential Pressure - Piezoresistive | |
|---|---|
| Measuring range | -200 to 200 hPa |
| Accuracy |
±0.5 hPa (-49.9 to 49.9 hPa) ±1.5 % of mv (Remaining Range) |
| Resolution | 0.1 hPa |
| Absolute Pressure | |
|---|---|
| Measuring range | +600 to +1150 hPa |
| Accuracy | ±10 hPa |
| Resolution | 1 hPa |
| Flue gas O₂ | |
|---|---|
| Measuring range | 0 to 25 Vol.% |
| Accuracy | ±0.2 Vol.% |
| Resolution | 0.01 Vol.% |
| Reaction time t₉₀ | < 20 s |
| Flue gas CO (with H₂-compensation) | |
|---|---|
| Measuring range | 0 to 10000 ppm |
| Accuracy |
±10 ppm or ±10 % of mv (0 to 200 ppm) ±20 ppm or ±5 % of mv (201 to 2000 ppm) ±10 % of mv (2001 to 10000 ppm) |
| Resolution | 1 ppm |
| Reaction time t₉₀ | < 40 s |
| Flue gas COlow | |
|---|---|
| Measuring range | 0 to 500 ppm |
| Accuracy |
±2 ppm (0 to 39.9 ppm) ±5 % of mv (Remaining Range) data corresponds to 20°C ambient temperature. Additional temperature coefficient 0.25% of reading/K. |
| Resolution | 0.1 ppm |
| Reaction time t₉₀ | < 40 s |
| Flue gas NO | |
|---|---|
| Measuring range | 0 to 4000 ppm |
| Accuracy |
±5 ppm (0 to 99 ppm) ±5 % of mv (100 to 1999 ppm) ±10 % of mv (2000 to 4000 ppm) |
| Resolution | 1 ppm |
| Reaction time t₉₀ | < 30 s |
| Flue gas Nolow | |
|---|---|
| Measuring range | 0 to 300 ppm |
| Accuracy |
±2 ppm (0 to 39.9 ppm) ±5 % of mv (Remaining Range) |
| Resolution | 0.1 ppm |
| Reaction time t₉₀ | < 30 s |
| Flue gas NO₂ | |
|---|---|
| Measuring range | 0 to 500 ppm |
| Accuracy |
±10 ppm (0 to 199 ppm) ±5 % of mv (Remaining Range) |
| Resolution | 0.1 ppm |
| Reaction time t₉₀ | < 40 s |
| Flue gas SO₂ | |
|---|---|
| Measuring range | 0 to 5000 ppm |
| Accuracy |
±10 ppm (0 to 99 ppm) ±10 % of mv (Remaining Range) |
| Resolution | 1 ppm |
| Reaction time t₉₀ | < 40 s |
| Flue gas Draught | |
|---|---|
| Measuring range | -40 to +40 hPa |
| Accuracy |
±0.03 hPa (-2.99 to +2.99 hPa) ±1.5 % of mv (Remaining Range) |
| Resolution | 0.01 hPa |
| Temperature | |
|---|---|
| Measuring range | -40 to +1200 °C |
| Accuracy |
±0.5 °C (0 to +99 °C) ±0.5 % of mv (Remaining Range) |
| Resolution | 0.1 °C |
| Flue gas degree of effectivity, Eta (calculated) | |
|---|---|
| Measuring range | 0 to 120 % |
| Resolution | 0.1 % |
| Flue gas loss (calculated) | |
|---|---|
| Measuring range | 0 to 99.9 % |
| Resolution | 0.1 % |
| Flue gas dewpoint (calculated) | |
|---|---|
| Measuring range | 0 to 99.9 °Ctd |
| Resolution | 0.1 °Ctd |
| Flue gas CO₂ calculation (calculated from O₂) | |
|---|---|
| Measuring range | 0 to CO₂ max |
| Accuracy | ±0.2 Vol.% |
| Resolution | 0.1 Vol.% |
| Reaction time t₉₀ | < 40 s |
| General technical data | |
|---|---|
| Dimensions | 283 x 103 x 65 mm |
| Operating temperature | -5 to +50 °C |
| Housing | TPE PC |
| Protection class | IP40 |
| Warranty | Measuring instrument 2 years (excepting wearing parts e.g. gas sensors, etc.); rechargeable battery: 12 months; CO sensor: 12 months; Sensor NOlow: 12 months; Sensor NO2: 12 months; Sensor SO₂: 12 months; Sensor O₂: 18 months; NO sensor 12 months; COlow sensor 12 months |
| Display size | 160 x 240 pixels |
| Display function | Graphic display |
| Power supply | Battery block 3.7 V / 2.4 Ah, Mains unit 6.3 V / 2 A |
| User defined fuels | 10 user-defined fuels incl. test gas as fuel |
| Max. neg. press./flue gas | -200 mbar |
| Max. pos. press./flue gas | 50 mbar |
| Hose length | max. 7.8 m (corresponds to two probe hose extensions) |
| Pump flow | 0.6 l/min (regulated) |
| Maximum | 100 folders |
| Storage per folder | Max. 10 sites |
| Storage per site | Max. 200 logs |
| Miscellaneous | The max. number of logs is determined by the number of folders or sites |
| Storage temperature | -20 to +50 °C |
| Weight | 960 g |